probe for measurement of luminance | LP-471-LUM 2
443,00 EUR
inkl. 19 % USt -> Preis zzgl. Versandkosten
Lieferzeit bzw. Fertigungszeit variiert - bitte erfragen!
Art.Nr.: LP-471-LUM 2Produktbeschreibung
general information:
Spectral response according to the photopic curve, angular field 2°.
Measuring range: 1.0 cd/m² to 2,000·10³ cd/m²
application areas:
The sensor measures luminance like a human eye, e.g.: monitors and lamps, etc. diaphanoscope, readout of X-ray plates, light radiation of PC monitors and reflections by white surfaces.
specifications:
measuring range and resolution (cd/m²): |
measuring range: 1.0...1,999.9 resolution: 0.1 measuring range: ...19,999 resolution: 1 measuring range: ...199.99·103 resolution: 0.01·103 measuring range: ...1,999.9·103 resolution: 0.1·103 |
|
optical angle: | 2° | |
spectral range: | in agreement with standard photopic curve V(λ) (p.r.t. data sheet) | |
α (temp. coefficient) f6 (T): | <0.05 % K | |
calibration uncertainty: | <5 % | |
f'1 (in agreement with photopic response V(λ)): | <8 % | |
f3 (linearity): | <1 % | |
f4 (instrument reading error): | <0.5 % | |
f5 (fatigue): | <0.5 % | |
class: | C | |
drift after 1 year: | <1 % | |
working temperature: | 0°C to +50°C | |
reference standards: | CIE n.69 - UNI 11142 |
catalogue page as .pdf: |
Empfohlenes Zubehör | Weitere Artikel
Photo- und Radiometer | HD-2302-0339,00 EUR |
|
Photo-Radiometer | HD21020,00 EUR |