Deutsch  English 


conductivity meter | LF9648

conductivity meter | LF9648


520,00 EUR

 inkl. 19 % USt  -> Preis zzgl. Versandkosten

Lieferzeit bzw. Fertigungszeit variiert - bitte erfragen!

Art.Nr.: LF9648
terminal block A:
1: input for 2- or 4-elektrode measuring cells; temperature comp. Pt100:
3: input 2- or 4-elektrode measuring cells; temperature comp. Pt1000:
terminal block B:
00: not set:
2R: 2 relay outputs (+ 88,70 EUR) :
2T: 2 electronic outputs (+ 88,70 EUR) :
terminal block C:
00: not set:
2R: 2 relay outputs (+ 88,70 EUR) :
2T: 2 electronic outputs (+ 88,70 EUR) :
AO: analog output 0/4 to 20 mA and 0/2 to 10 V DC (+ 158,50 EUR) :
2A: 2 analog outputs 4 to 20 mA, passive (+ 211,00 EUR) :
terminal block D, auxiliary voltage:
0: 230 V AC ±10 % 50-60 Hz:
1: 115 V AC ±10 % 50-60 Hz (+ 17,80 EUR) :
4: 24 V AC ±10 % 50-60 Hz (+ 17,80 EUR) :
5: 24 V DC ±15 % (+ 35,20 EUR) :
options MA:
01: min and max value memory (+ 52,80 EUR) :
14: measurement and monitoring acc. to USP<645> (+ 105,90 EUR) :
unit (imprint in the unit field) :

Bitte beachten Sie stets die Bestellhinweise auf den einzelnen Datenblättern!

Produktbeschreibung


general information:

The conductivity meter LF9648 has been designed for the measurement of the condutivity in liquids with conductive measuring cells. Depending on the conductivity of the medium, 2-electrode measuring cells (e.g. ultrapure water) or 4-electrode measuring cells (e.g. process water, detergent solutions, acids, alkalis, etc.) are required.

specifications:

auxiliary energy

auxiliary voltage: 230 V AC ±10 %; 115 V AC ±10 %;
24 V AC
±10 % or 24 V DC ±15 %
power consumption: max. 3.5 VA, with analog output 5 VA
operating temperature: -10°C to +55°C
CE conformity:EN 61326-1:2013, EN 60664-1:2007


input
MR conductivity:0...2.000(0) μS/cm to
0...
2,000/200(0) mS/cm (at +25°C)
cell constant:0.080 to 9.999
basic accuracy:0.5 % of measured value, ±2 digit
temperature comp.:non linear for ultrapure water and natural water or linear programmable from 0.000 %/K to 9.999 %/K
MR temperature:-50.0°C to +200.0°C;
sensor Pt100 or Pt1000
basic accuracy:±0.2°C


display
display:LED red, 14.2 mm
display range:2,000(0) digit with leading zero suppression
additional display:LED 2-digit red, 7 mm
(indication of parameter and switching state)


outputs
relay:SPDT <250 V AC <250 VA <2 A,
 <300 V DC <50 W <2 A
transistor:max. 35 V AC/DC, 100 mA,
with electronic current limit
analog output:active: 0/4 mA to 20 mA burden ≤500 Ω;
0/2 V to 10 V burden >500 Ω, galv. isolated
output switches automatically (burden dependent)
passive: 4 mA to 20 mA, ext.
burden = RA[Ω]≤ (supply - 5 V) ÷ 0.02 A;
auxiliary voltage 5 V DC to 30 V DC
accuracy: 0.1 %; TK 0.01 %/K


housing
housing:panel mounting DIN 96 x 48,
material PA6-GF; UL94V-0
dimensions:front 96 x 48 mm, mounting depth 100 mm
weight:max. 390 g
connection:clamp terminals, 0.08 mm2 to 1.5 mm2,
AWG28..AWG14





For further details p.r.t. data sheet:




connection diagrams:





instruction manual as .pdf:


 

 

 

Empfohlenes Zubehör | Weitere Artikel

Verbindungskabel | SKM8E

Verbindungskabel | SKM8E

75,90 EUR

Verbindungskabel | SKM4E

Verbindungskabel | SKM4E

58,90 EUR

Warenkorb

Sie haben noch keine Artikel in Ihrem Warenkorb.

Warenkorb »

Herzlich willkommen

E-Mail-Adresse: