Deutsch  English 


conductivity converter | UNICON®-LF

conductivity converter | UNICON®-LF


862,00 EUR

 inkl. 19 % USt  -> Preis zzgl. Versandkosten

Lieferzeit bzw. Fertigungszeit variiert - bitte erfragen!

Art.Nr.: UNICON-LF
construction type:
1: output 4 to 20 mA for conductivity, 2 contactless alarm outputs:
2: as 1, but 2nd measuring range for conductivity, output 4 to 20 mA for temperature (+ 96,80 EUR) :
type of installation:
01: head mounting, directly on measuring cell:
02: field mounting with connection socket brass nickel-plated (+ 96,80 EUR) :
03: field mounting with connection socket VA 1.4571 (+ 105,90 EUR) :
measuring principle:
4: 4-electrode measurement (2-electrode measuring cells connectable):
temperature compensation:
1: Pt100:
3: Pt1000:
options MA:
14: measurement and monitoring acc. to USP<645> (+ 105,90 EUR) :

Bitte beachten Sie stets die Bestellhinweise auf den einzelnen Datenblättern!

Produktbeschreibung


general information:

The conductivity converter UNICON®-LF
has been designed for the measurement of the condutivity in liquids. Depending on the conductivity of the medium, 2-electrode measuring cells (e.g. ultrapure water) or 4-electrode measuring cells (e.g. process water, detergent solutions, acids, alkalis, etc.) are required.


specifications:


auxiliary voltage

loop voltage:
14 V DC to 30 V DC, 2-wire connection
operating temperature: 0°C to +50°C
CE conformity: EN 61326-1:2013


conductivity output

output signal: 4 mA to 20 mA
unit:
progammable μS/cm; mS/cm; kΩ/cm; MΩ/cm
decimal places:
0 to 3 digits (depending on unit)
display range:
500 to 9,999 digit (depending on unit and decimal places)
min./max. measuring range:
0..5.00 μS/cm to 0..500.0 mS/cm;
0..0.500 μS/cm / 0..50.0 μS/cm
with measuring cell for ultrapure water

temperature comp:
non linear for ultrapure water and
natural water or linear programmable from
0.000
%/°C to 8.000 %/°C
cell constant:programmable in the range 0.080 to 4.000
standard error:±0.5 % of measured value ±2 digit


temperature output
output signal:4 mA to 20 mA
burden:RA ≤ (UB-14 V) ÷ 0,02 A
temperature probe:Pt100 or Pt1000 acc. to DIN IEC 751
unit:°C, °F programmable
measuring range:-40.0°C to +160.0°C


alarm outputs
transistor:14 V DC to 30 V DC, max. 60 mA
voltage drop:<2 V


measuring range switch over
Ri:>10 kΩ
MB1 active:U = 0 V DC to 3 V DC
MB2 active:U = 12 V DC to 30 V DC


display
display:LCD dot matrix, 3.8 mm character height
range:2 lines 16 characters each


housing
housing:head case / field case
material:housing polyamide glass fibre reinforced
PA6-GF/GK 15/15, front foil polyester
dimensions:100 x 100 x 60 mm (W x H x D)
weight:max. 360 g
connection:screw terminal with wire protection,
2.5 mm2 flexible, 4 mm2 wire and
plug-in cable for sensor
ingress protection:IP65, terminals IP20 acc. to BGV A3




For further details p.r.t. data sheet:



instruction manual as .pdf:

 


 

 

Empfohlenes Zubehör | Weitere Artikel

Verbindungskabel | SKM8

Verbindungskabel | SKM8

99,20 EUR

Verbindungskabel | SKM4B

Verbindungskabel | SKM4B

75,90 EUR

Warenkorb

Sie haben noch keine Artikel in Ihrem Warenkorb.

Warenkorb »

Herzlich willkommen

E-Mail-Adresse:
Passwort:
Passwort vergessen?

Top Artikel