complete set | GMH 3151-WPD5-GKK31
560,00 EUR
incl. 19 % VAT excl. shipping costs
delivery time usually 0-4 working days for standard construction type
Product.Nr.: GMH3151-WPD5-GKK31Product Description
set GMH 3151-WPD5-GKK31 consisting of:
- 4½-digit display, probes with higher resolution up on request
- peak value memory 1,000 measurments / s
- analog output 0 V to 1 V
- digitally adjustable
- integrated alarm horn
max. display range: | -19,999 to +19,999 digit |
measuring range: | corresponding to probe |
overload: | corresponding to probe |
resolution: | corresponding to probe |
accuracy (device): | ±0.1 % FS ±1 digit (at nominal temperature = +25°C) |
pressure units: | mbar, bar, Pa, kPa, MPa, mmHg, PSI, mH2O, can be selected via keystroke (The choice of a specific pressure unit is only possible, if its whole measuring range is displayable on the display of the device and the sensor is supporting this resolution.) |
probe connection: | 1 sensor socket 6-pole shielded Mini-DIN socket with locking for GMSD/MSD-sensors, automatic probe detection and appropriate setting of measuring range |
display: | 2x 4½-digit LCD |
output: |
interface or AAG |
serial interface: | direct connection to RS232 or USB interface of a PC via electrically isolated interface converter GRS 3100, GRS 3105 or USB 3100 N |
analog output: | 0 V to 1 V, freely adjustable (resolution 12 bit) |
power supply: | 9 V battery, d.c. connector for external 10.5 V to 12 V direct voltage supply (suitable power supply: GNG10/3000) |
sensor adjustment: | digital offset and scale input |
measuring cycle | "slow": 4 measurements / s "fast" / "peak-detect": ≥1,000 measurements / s |
peak-detect: | the min- / max- value memory stores unfiltered pressure peaks ≥1 ms |
logger functions: | manual data sets: 99 data sets cycle data sets: 10,000 adjustable cycle time: 1 s to 3,600 s (max. 64 recording sequences), the logger is started or stopped via keypad or interface, the software GSOFT3050 (see accessories) is available for comfortable read-out of logger data |
averaging function: | integrates the measured values during a selectable period of time and then calculates the average display value |
sea-level-correction: | when connecting an abs. pressure probe the barom. air pressure can also be displayed corrected to sea level „zero“ (air pressure compensation is achieved by entering the meters above sea level „zero“) |
battery life: | approx. 180 h (4 measurements / s) approx. 40 h (1000 measurements / s) |
operating conditions: | -25°C to +50°C, 0 % to +95 % RH (non condensing) |
housing: | impact-resistant ABS plastic housing, membrane keypad, transparent panel GMH 315x-EX only: integrated pop-up clip |
dimensions: | 142 x 71 x 26 mm (H x W x D) |
weight: | approx. 150 g |
catalogue page calibration certificates as .pdf: