data logger | HD35ED-G-14BN-TVI-E
609,00 EUR
incl. 19 % VAT excl. shipping costs
delivery time processing time as may be necessary - please ask!
Product.Nr.: HD35ED-G-14BN-TVI-EProduct Description
wireless data logger for temperature, humidity and atmospheric pressure with fixed vertical T-/humidity probe
general information:
HD35ED-G-14BN-TVI-E stores the measured values in its internal memory (22,000 samples) and transmits the acquired data to the base unit automatically at regular intervals or upon request. Fixed vertical probe for temperature and humidity with temperature sensor integrated in the humidity module. Calculated measurement quantities: dew point, wet bulb temperature, absolute humidity, mixing ratio, partial vapour pressure. Acoustic alarm with internal buzzer. Configuration via HD35-AP-S software or front keyboard. Power supply by internal battery. Installation with fixed or removable wall mount (supplied mount or optional flanges).
specifications:
humidity (r.h.): |
measuring principle: | capacitive |
measurement range: | 0 % to 100 % r.h. |
|
resolution: |
0.1 % r.h. |
|
accuracy (at +23°C): | ± 1.8 % r.h (0 % to 80 % r.h.) ± [1.8 + 0.11 * (r.h. -80)] % r.h. (remaining range) | |
working temperature: | -40°C to +105°C (r.h. max = [100-2 * (T-80)] at T = +80°C to +105°C) |
|
temperature T: |
sensor: | sensor integrated in humidity module |
measurement range: | -40°C to +105°C |
|
resolution: |
0.1°C |
|
accuracy: | ±0.2°C in the range 0°C to +60°C ±(0.2 – 0.05 * T)°C in the range T = -40°C to 0°C ±[0.2 + 0.032 * (T-60)]°C in the range T = +60°C to +105°C | |
barometric air pressure (Patm): |
measuring principle: | piezo-resistive |
measurement range: | 300 hPa to 1,100 hPa |
|
resolution: |
0.1 hPa |
|
accuracy: | ±0.5 hPa (800 hPa to 1,100 hPa) at T = +25°C ±1 hPa (300 hPa to 1,100 hPa) at T = 0°C to +50°C | |
frequency of transmitting: | 868 MHz | |
transmitting range: | 300 m (E, J) / 180 m (U) in open field (can be reduced due to the presence of obstacles or adverse atmospheric conditions) | |
logging interval: | 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30, 60 min | |
working temperature: |
-20°C to +70°C | |
relative working humidity: | 0 % to 85 % r. F., non-condensing | |
ingress protection: | IP64 | |
power supply: | internal, non-rechargeable lithium thionyl choride (Li-SOCl2) battery, 3.6 V, format 'AA', 2-pole Molex 5264 connector | |
battery life: | 2 years typical (without repeaters, measurement interval 10 s and logging interval 30 s) |
catalogue page as .pdf: |
Recommended products
repeater | HD35-RE-E324,00 EUR |
|
interface | HD35-AP-D-E156,50 EUR |
|
interface | HD35-AP-W-E521,00 EUR |
|
interface | HD35-AP-G-E498,00 EUR |
|
non-rechargeable 3.6 V lithium-thionyl chloride battery | HD35-BAT221,20 EUR |
|
alarm module | HD35-ED-ALM-E319,00 EUR |
|
software | HD35-AP-S68,30 EUR |
|
web data logger | HD 50...0,00 EUR |